The WST-1 assay is the most common test to assess the in vitro cytotoxicity of chemicals. Tetrazolium-based assays can, however, be affected by the interference of tested chemicals, including carbon nanotubes or Mg particles. Here, we report a new interference of Mn materials with the WST-1 assay. Endothelial cells exposed to Mn particles (Mn alone or Fe-Mn alloy from 50 to 1600 μg/ml) were severely damaged according to the WST-1 assay, but not the ATP content assay. Subsequent experiments revealed that Mn particles interfere with the reduction of the tetrazolium salt to formazan. Therefore, the WST-1 assay is not suitable to evaluate the in vitro cytotoxicity of Mn-containing materials, and luminescence-based assays such as CellTiter-Glo® appear more appropriate.