The AVI was classified as normal, too long or too short.

Download (0 kB)
figure
posted on 23.02.2015 by Wolfram C. Poller, Henryk Dreger, Marius Schwerg, Christoph Melzer

To demonstrate the three categories of transmitral inflow pattern, one patient was analyzed with an optimal, a too short and a too long AV-sense interval. A) A normal AVI (135 ms) presents with separated E- and A waves, the A wave maximized in size and length and no diastolic mitral regurgitation. B) When the AVI is programmed too short (70 ms) an A wave truncation occurs (short, small and abruptly terminating A wave). C) A too long AVI (250 ms) presents with a complete E/A fusion. Such E/A fusion occurred in 19.4% (CI 12.6–26.2%) of the analyzed patients in sinus rhythm.